• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; INTERNA UNIVERSAL PLATFORM IMPLANT CA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; INTERNA UNIVERSAL PLATFORM IMPLANT CA Back to Search Results
Model Number IIPUCA4010
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Pain (1994); Bruxism (4427)
Date of Event 10/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician indicates that he placed the implant on (b)(6) 2020 and two weeks later it was noted that the implant had not osseointegrated.Patient with bone quality ii.In the event the patient experienced: infection and inflammation.
 
Additional Manufacturer Narrative
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
INTERNA UNIVERSAL PLATFORM IMPLANT CA
Manufacturer (Section D)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street, 14
minano menor, alava 01510
SP  01510
Manufacturer (Section G)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street, 14
minano menor, alava 01510
SP   01510
Manufacturer Contact
jose rivero
leonardo da vinci street, 14
minano menor, alava 01510
SP   01510
MDR Report Key19689439
Report Number3004417597-2024-00019
Device Sequence Number15224104
Product Code DZE
UDI-Device Identifier08435389812203
UDI-Public08435389812203
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K151391
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date11/01/2024
Device Model NumberIIPUCA4010
Device Lot NumberAI05332.Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/15/2021
Initial Report FDA Received Date07/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-