| Model Number |
PCF-H190TI |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the colonovideoscope exhibited foreign materials in the air/water tube and clogged nozzle.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.The legal manufacturer was unable to confirm whether the customer's reprocessing steps were deviated from the instructions for use.Based on the results of the investigation, olympus confirmed foreign material remained in the device, but the type of the material or root cause cannot be identified.Physical damage was not found at the locations.The event can be detected/prevented by following the instructions for use: instructions for use states the detection method in gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection.Instructions for use states the preventive measure in gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|
| |
|