| Model Number |
CCWTT0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Disturbances (2140)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported following an intraocular lens (iol) implant procedure, the patient experienced glare and reading was very difficult.The lens was explanted and replaced with another lens in a secondary procedure.The clinical reason for the explant was dysphotopsias secondary to iol.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100percentage assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the company model, 19.5 diopter (add power plus2.2 or plus3.2) lens was replaced with an unspecified atiol model due to reported dysphotopsia.The product has not been received to evaluate.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|