The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the foley catheter balloon just deflated, and it did not fall out so they were not aware that the patient was not able to void and had ended up in the hospital with urosepsis.It was noted that this happened with 2 foleys with same lot numbers.Per customer on 20jun2024, it was reported that in february, patient experienced a deflated balloon but was unaware because it did not fall out.Customer noticed that they had not voided in 9 hours and the catheter was not draining.Customer was hospitalized on (b)(6) 2024 and was diagnosed with uro-sepsis.Medical intervention was reported and customer received iv antibiotics to treat the infection.The second occurrence happened on 11jun2024 with the same lot# nghv2411 which was reported for both events.
|
|
It was reported that the foley catheter balloon just deflated, and it did not fall out, so they were not aware that the patient was not able to void and had ended up in the hospital with urosepsis.It was noted that this happened with 2 foleys with same lot numbers.Per customer on (b)(6)2024, it was reported that in (b)(6) patient experienced a deflated balloon but was unaware because it did not fall out.Customer noticed that he had not voided in 9 hours and the catheter was not draining.Customer was hospitalized on (b)(6) 2024 and was diagnosed with uro-sepsis.Medical intervention was reported and customer received iv antibiotics to treat the infection.The second occurrence happened on (b)(6)2024 with the same lot# nghv2411 which was reported for both events.
|
|
The reported event was inconclusive.The conditions of the sample did not allow further evaluation.Received one-photo sample of the lubrisil silicone foley box with product information.The reported failure could not be evaluated therefore this investigation is considered inconclusive.Root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be imperfection in balloon due to process.Dhr review did not show any problems or conditions.The instructions for use were found adequate and state the following: 1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard ez-lok sampling port with antiseptic wipe.(fig.2) 3.Using aseptic technique, position the syringe in the center of the sampling port.Press the syringe firmly and twist gently to access the sampling port.(fig.3) 4.Slowly aspirate urine sample into syringe and remove syringe from sample port.5.Unkink tubing if necessary and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab.Sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter insertion perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance secure the foley catheter, use the statlock foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate.Leave foley catheter in place only as long as needed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|