Catalog Number MB3796
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event or Problem Description
It was reported that a cartridge would not release from the mobi-c implant in the disc space, and the doctor had to use a needle driver to twist each cartridge out and dislodge the device.Another implant was opened and used to complete the procedure.There was no patient impact.
 
Additional Manufacturer Narrative
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in g1 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed no signs of damage.The implant was attached to an inserter (mb9001r-1 lot 310232903/20) and was able to be implanted into a cervical spine model.The implant was able to release from the cartridge as expected into the disc space.Root cause: a definitive root cause cannot be determined with the information provided.No device problem was found.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that a cartridge would not release from the mobi-c implant in the disc space, and the doctor had to use a needle driver to twist each cartridge out and dislodge the device.Another implant was opened and used to complete the procedure.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBI-C CERVICAL DISC, 17 X 19, H6
Common Device Name
MOBI-C CERVICAL DISC PROTHESIS
Manufacturer (Section D)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR  10300
Manufacturer (Section G)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR   10300
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key19691293
Report Number3004788213-2024-00054
Device Sequence Number18173313
Product Code MJO
UDI-Device Identifier03662663002059
UDI-Public(01)03662663002059(17)250201(10)5345911
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P110009
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Catalogue NumberMB3796
Device Lot Number5345911
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/17/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/08/2024
Supplement Report FDA Received Date12/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
Patient Weight113 KG
Patient EthnicityNon Hispanic
-
-