• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC HDO; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC HDO; INTRAOCULAR LENS Back to Search Results
Lot Number F223557
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Eye Pain (4467)
Date of Event 05/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Lenstec received an email indicating "the patient vision was 0.4 on the second day after surgery.".
 
Additional Manufacturer Narrative
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.The device reported remains implanted and therefore, not returned to lenstec.All endotoxin testing met specifications.There is no evidence to suggest that any aspect of this device was responsible for the reported inflammation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HDO
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
Manufacturer (Section G)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB  
Manufacturer Contact
jimmy chacko
1765 commerce ave. n.
st. petersburg, FL 33716
7275712272
MDR Report Key19692030
Report Number9613160-2024-00045
Device Sequence Number12769417
Product Code HQL
UDI-Device Identifier00844369037471
UDI-Public00844369037471
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Lot NumberF223557
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/27/2024
Initial Report FDA Received Date07/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-