This supplemental report is being submitted to provide the legal manufacturer¿s final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than a year since the subject device was manufactured.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Additionally, the inspection of the broken part of the knife wire found a reportable malfunction of the tearing of the wire coating.The broken part was burnt and melted.No other abnormalities that could lead to breakage of the knife wire were confirmed.Based on the results of the investigation, a definitive root cause could not be established.A likely mechanism causing the cutting wire breakage might be the following.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.Under circumstances described above 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire melted.When the cutting wire was tightened, a load was applied to the melted portion of the cutting wire.This caused the cutting wire to break.A likely mechanism causing the cutting wire breakage might be the following: it is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.The event can be detected/prevented by following the applicable chapters in the instructions for use which state: since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Olympus will continue to monitor field performance for this device.
|