| Catalog Number |
3MAXC |
| Medical Device Problem Codes |
Fracture (1260); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure in the left internal carotid artery (ica) terminus using a penumbra system red 68 reperfusion catheter (red68), a penumbra system 3max reperfusion catheter (3maxc), a non-penumbra sheath, a stent retriever, and a guidewire.It was noted that the patient''s anatomy was tortuous.During the procedure, the physician completed passes in the target location using the red68 and 3maxc.The physician then re-advanced the red68 and 3maxc to the clot and attempted to advance the stent retriever through the red68.It was reported that resistance was encountered when placing the red68 and stent retriever due to the patient¿s anatomy, causing the physician to use additional force.After several attempts to advance the stent retriever, the physician decided to remove it with the red68 and 3maxc and again experienced resistance.Upon removal, it was noted that both the red68 and the 3maxc fractured into two pieces approximately five centimeters from the proximal end.Therefore, the red68 and the 3maxc were not used for the remainder of the procedure.The procedure was completed using a non-penumbra aspiration catheter and the same sheath.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 1.3005168196-2024-00238.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned 3maxc revealed that the catheter was fractured, and the inner coil winds were exposed.Stress marks were present near the fracture, this indicated that the device was likely kinked prior to the fracture.It was reported that the patient¿s anatomy was tortuous.If the 3maxc is manipulated through tortuous anatomy, damage such as a kink and subsequent fracture may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2024-00238, h3 other text : placeholder.
|
| |
|
Search Alerts/Recalls
|