| Model Number |
KNOTLESS DIS CLAV PLATE BUTTON TIGHTROP |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 6/17/2024, it was reported by a sales representative via phone that an ar-2658tr knotless distal clavicle plate button tightrope had the button come off the inserter after drilling with an ar-2272 drill pin button 3.7 mm through the clavicle.Suture and button were removed from the patient.Then, an ar-2372blo knotless ac tightrope open repair implant was used, and it prematurely came off the inserter.It was cut out, and suture and button were removed from the patient.The case was completed using an ar-2658tr knotless distal clavicle plate button tightrope with no issues.The case was delayed 45 minutes.It is unknown if additional anesthesia was administered to the patient.No adverse effects on the patient were reported due to the issue.This was discovered during a distal clavicle fracture procedure on (b)(6) 2024, with no reported patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Per surgical technique lt1-02553-e 3-insert the coracoid button into the clavicle tunnel.Light taps on the inserter handle can aid in advancing the button through the tunnels.4-insert the coracoid button through the clavicle and the coracoid tunnel.Fluoroscopy may be used to confirm proper button alignment and that it extends below the inferior cortex of the coracoid.The most likely cause for the reported failure can be attributed to user error of the device due to misalignment insertion, and/or prying/leveraging of the device during insertion.
|
| |
|
Search Alerts/Recalls
|