• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. KNOTLESS DIS CLAV PLATE BUTTON TIGHTROP; BONE FIXATION PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. KNOTLESS DIS CLAV PLATE BUTTON TIGHTROP; BONE FIXATION PLATE Back to Search Results
Model Number KNOTLESS DIS CLAV PLATE BUTTON TIGHTROP
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 6/17/2024, it was reported by a sales representative via phone that an ar-2658tr knotless distal clavicle plate button tightrope had the button come off the inserter after drilling with an ar-2272 drill pin button 3.7 mm through the clavicle.Suture and button were removed from the patient.Then, an ar-2372blo knotless ac tightrope open repair implant was used, and it prematurely came off the inserter.It was cut out, and suture and button were removed from the patient.The case was completed using an ar-2658tr knotless distal clavicle plate button tightrope with no issues.The case was delayed 45 minutes.It is unknown if additional anesthesia was administered to the patient.No adverse effects on the patient were reported due to the issue.This was discovered during a distal clavicle fracture procedure on (b)(6) 2024, with no reported patient harm.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Per surgical technique lt1-02553-e 3-insert the coracoid button into the clavicle tunnel.Light taps on the inserter handle can aid in advancing the button through the tunnels.4-insert the coracoid button through the clavicle and the coracoid tunnel.Fluoroscopy may be used to confirm proper button alignment and that it extends below the inferior cortex of the coracoid.The most likely cause for the reported failure can be attributed to user error of the device due to misalignment insertion, and/or prying/leveraging of the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNOTLESS DIS CLAV PLATE BUTTON TIGHTROP
Common Device Name
BONE FIXATION PLATE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19700094
Report Number1220246-2024-06652
Device Sequence Number7655573
Product Code HRS
UDI-Device Identifier00888867271814
UDI-Public00888867271814
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K143139
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKNOTLESS DIS CLAV PLATE BUTTON TIGHTROP
Device Catalogue NumberAR-2658TR
Device Lot Number15204514
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2024
Supplement Date Received by Manufacturer03/17/2025
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date04/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-