Additional information was provided in the product was not returned.Complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.The ifu (instructions for use) instructs: during device preparation and implantation of the company iol (intraocular lens) with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Lens may become stuck in the device: if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to become ¿stuck¿ in the device due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|