• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
A health care professional reported that during an intraocular lens (iol) implant procedure, lens misfired and did not come out the injector properly at time of surgery.Additional information was requested.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in the product was not returned.Complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.The ifu (instructions for use) instructs: during device preparation and implantation of the company iol (intraocular lens) with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Lens may become stuck in the device: if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to become ¿stuck¿ in the device due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19701122
Report Number1119421-2024-01291
Device Sequence Number15367507
Product Code KYB
UDI-Device Identifier00380652394888
UDI-Public00380652394888
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAU00T0
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/11/2024
Supplement Date Received by Manufacturer08/22/2024
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date09/19/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-