• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. AUTOCLAVABLE CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number OTV-S7H-NA-12E
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the subject device had a picture that was not clear.The issue was found during an unknown procedure.There were no reports of patient harm.
 
Additional Manufacturer Narrative
E1 - address 1: (b)(6).E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: images are not displayed, cable flooding and cable damage.Based on the results of the investigation, it is likely the following led to the malfunction: cause linked to device but unable to trace more specifically olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that the subject device had a picture that was not clear.The issue was found during an unknown procedure.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCLAVABLE CAMERA HEAD
Common Device Name
CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19701217
Report Number3002808148-2024-34636
Device Sequence Number14365888
Product Code FET
UDI-Device Identifier04953170141867
UDI-Public04953170141867
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K955404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-S7H-NA-12E
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/24/2024
Supplement Date Received by Manufacturer07/23/2024
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date07/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-