• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030457
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
Date of Event 06/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: egiaustnd, egiaustnd endogia ultra univ std stap, (lot#p2c0582s).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, in an open radical pancreaticoduodenectomy, when detaching the duodenum, the staples did not form a good b-shape and the anastomosis was bleeding.The staple line was also incomplete centrally.The staple line was manually sutured to fix the issue.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the loading unit was partially fired to the 4cm cut line.The anvil was bowed.The clamping mechanism was deformed.The device had a buckled knife bar assembly.It was reported that the staples did not form properly and the staple line was incomplete.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if an application over tissue that is beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In any of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: select a loading unit with the appropriate staple size for the tissue thickness.Overly thick or thin tissue may result in unacceptable staple formation.Always include the combined thickness of the tissue and of any staple line reinforcement material in use when choosing the proper staple cartridge.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19701888
Report Number1219930-2024-03104
Device Sequence Number7991521
Product Code GDW
UDI-Device Identifier10884523002881
UDI-Public10884523002881
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number030457
Device Catalogue Number030457
Device Lot NumberT2F056X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2024
Supplement Date Received by Manufacturer08/20/2024
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date09/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient Age69 YR
Patient SexMale
Patient Weight72 KG
-
-