Model Number Ø10X 30MM BC IF SCRW, VENTED
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2024 it was reported by a sales representative via phone that an ar-4030c-10 fastthread biocomposite interference screw cracked during insertion.All fragments were retrieved from the patient.There was a minute delay in the procedure.The case was completed using another ar-4030c-10 fastthread biocomposite interference screw with no issues.This was discovered during a acl repair procedure on (b)(6) 2024, with no reported patient harm.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Per dfu-0111.Precautions: 6.Bio-absorbable interference screw only: it is important to completely seat the screwdriver to prevent potential stripping of the hex and/or screw fracture during insertion or removal.The most likely cause for the reported failure can be a user error due to misaligned insertion, and/or prying/levering the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ø10X 30MM BC IF SCRW, VENTED
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19702559
Report Number1220246-2024-06669
Device Sequence Number15397813
Product Code MAI
UDI-Device Identifier00888867277670
UDI-Public00888867277670
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K202535
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberØ10X 30MM BC IF SCRW, VENTED
Device Catalogue NumberAR-4030C-10
Device Lot Number15203218
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2024
Supplement Date Received by Manufacturer05/12/2025
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date05/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-