SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74121142
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 03/21/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on literature review ""clinical and radiographic outcomes of the birmingham hip resurfacing arthroplasty at a minimum follow-up of 10 years: results from an independent centre"", 1 (one) patient suffered from a varus femoral component shift and required a revision surgery 18 months after the primary metal-on-metal hip resurfacing procedure.During the revision surgery, the primary system was explanted and replaced by an unspecified uncemented tha metal-on-polyethylene system.The primary surgery was a bhr metal on metal due to osteoarthritis in congenital hip dysplasia.The outcome of the patient is unknown.No further information is available.
 
Additional Manufacturer Narrative
Literature citation: moroni a, miscione mt, orsini r, micera g, mosca s, sinapi f, de girolamo l, banfi g.Clinical and radiographic outcomes of the birmingham hip resurfacing arthroplasty at a minimum follow-up of 10 years: results from an independent centre.Hip int.2017 mar 31;27(2):134-139.Doi: 10.5301/hipint.5000424.Epub 2017 mar 21.Pmid: 28362050.H10: internal reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from literature sources that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
Section h10: it was reported that on literature review "clinical and radiographic outcomes of the birmingham hip resurfacing arthroplasty at a minimum follow-up of 10 years: results from an independent centre", 1 (one) patient suffered from a varus femoral component shift and required a revision surgery 18 months after the primary metal-on-metal hip resurfacing procedure.During the revision surgery, the primary system was explanted and replaced by an unspecified uncemented total hip arthroplasty metal-on-polyethylene system.The outcome of the patient is unknown.No further information is available.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the alleged devices.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.No other similar complaints have been identified for the head.This will continue to be monitored via routine trending, however it should be noted that this device is no longer sold.As no device batch numbers were provided for investigation, manufacturing record review could not be performed.If more information is received, this investigation will be reopened.The review of the most recent ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.To the date, no clinically sufficient data was provided to perform a medical investigation.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 42MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19705993
Report Number3005975929-2024-00112
Device Sequence Number7751639
Product Code NXT
UDI-Device Identifier03596010502773
UDI-Public3596010502773
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Literature
Initial Reporter Occupation Physician
Remedial Action Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 08/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number74121142
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2024
Supplement Date Received by Manufacturer08/14/2024
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2745-2015
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN BIRMINGHAM HIP RESURFACING (BHR) CUP
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Required Intervention; Hospitalization;
Patient Age52 YR
Patient SexFemale
-
-