• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX PRO-VENT ARTERIAL BLOOD SAMPLING KIT; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX PRO-VENT ARTERIAL BLOOD SAMPLING KIT; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number G1469J
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.Date of event: month and year of event have been provided, but day is unknown.E1.Initial reporter last name unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that with the device leakage of blood was observed during syringe use.There was no disinfection or sterilization, and no infectious disease occurred.There was patient involvement and no patient harm reported.
 
Additional Manufacturer Narrative
One decontaminated device was returned for evaluation.Visual examination of the returned sample showed a hole near the syringe shoulder and the luer slip was chipped.The noted condition was such that it was not possible to conduct a leak test.Based on the results of this investigation the complaint was confirmed, however it could not be determined how the damage to the syringe barrel occurred, therefore no root cause was established.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX PRO-VENT ARTERIAL BLOOD SAMPLING KIT
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr
keene NH 03431
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key19706613
Report Number1217052-2024-00057
Device Sequence Number6578653
Product Code JKA
UDI-Device Identifier00351688060238
UDI-Public(01)00351688060238(17)260403(10)4388757
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K952516
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/03/2026
Device Catalogue NumberG1469J
Device Lot Number4388757
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/07/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2024
Supplement Date Received by Manufacturer08/05/2024
Initial Report FDA Received Date07/09/2024
Supplement Report FDA Received Date08/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-