| Model Number |
B12STF |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic gastroplasty procedure, at the beginning, the seal tore, causing the loss of pneumoperitoneum.Another trocar was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that after removal of the obturator, the duckbill seal was noted to be stretched out.The seals failed an air leak test.The product was shipped back with the obturator inserted in the cannula and therefore the length of time it was inserted during shipment may have also resulted in the stretched seals.The bladed obturator when actuated cut cleanly through test media.The trocar was broken open to remove the circular seal to check for subassembly overall height.It was reported that there was seal damaged.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|