H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.A visual inspection of internal components revealed sheared teeth on the firing rack.The instrument was successfully loaded with a representative device.During the firing cycle, a skip was audible in the firing stroke.An access hole was cut into the instrument body for visualization of the firing rack.It was reported that a gear was skipping.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The sheared teeth on the firing rack may occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.Attempting to fire a reload that is in interlock.In each circumstance, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
According to the reporter, during a laparoscopic thoracoscopic segmentectomy, after assembling the third reload, it felt dull when p ulling the handle.There was an obvious stuck when closing the jaws.It was noted that the first two reload fired normally.A new handle was used to resolve the issue.No patient injury.Medtronic's initial evaluation of the incident is that a visual inspection of internal components revealed sheared teeth on the firing rack.
|