This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d9, g3, g6, h2, h3, h4, h6, h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Additionally, the following were found during inspection: flooding of image capture, flooding of cables, discoloration of electrical contacts and discoloration of main body.Based on the results of the investigation, it is likely the following led to the malfunction: a definitive root cause could not be identified.Olympus will continue to monitor field performance for this device.
|