Model Number VLP-2006
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Sterile samples from the reported lot were not returned for visual review and testing.The actual devices or magnified photos were not returned for review.If samples or photos become available at a later time, they will be reviewed and tested and the results will be included in the file.A follow-up report will be submitted at that time.A potential root cause for a needle detaching when slightly tugged during a procedure could be that the suture was not fully inserted within the end of the needle during the manufacturing process.It¿s also possible the device(s) are being grasped on or near the swaged end, damaging the suture, allowing it to break free from the needle component.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point¿.Without reviewing and receiving the actual reported device(s), receiving magnified photos of the actual devices, or receiving detailed information regarding the method utilized to remove the device from the packaging, preoperative preparation of the device, tools utilized to grasp the devices, details of the procedure performed, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
 
Event or Problem Description
It was reported during the surgical procedure the needle detached and was removed immediately.No patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUILL
Common Device Name
2-0 UNI PDO 20CM
Manufacturer (Section D)
SURGICAL SPECIALTIES CORPORATION
corredor tijuana rosarito 2000
#24702 b
tijuana baja california 22235
MX  22235
Manufacturer (Section G)
SURGICAL SPECIALTIES CORPORATION
1100 berkshire blvd.
ste 308
reading PA 19608
Manufacturer Contact
kelly knappenberger
corredor tijuana-rosarito 2000 #24702b
ejido francisco villa
tijuana 22235
MX   22235
MDR Report Key19715745
Report Number3010692967-2024-00050
Device Sequence Number8725237
Product Code NEW
UDI-Device Identifier00848782020751
UDI-Public00848782020751
Combination Product (Y/N)N
PMA/510(K) Number
K132268
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVLP-2006
Device Lot NumberD403AVR
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/10/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-