Catalog Number 730M0060
Medical Device Problem Codes Fracture (1260); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that two vitality cannulated drivers and a vitality t27 closure top inserter fractured intra-operatively while driving a screw, but that these fractures resulted in no procedural delays or impact to the patient.It was reported that the tips had become twisted and compromised over use and time.No additional instrumentation was required to complete the surgery and all pieces were removed prior to rod insertion.This is report one of three for this event.
 
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event or Problem Description
It was reported that two vitality cannulated drivers and a vitality t27 closure top inserter fractured intra-operatively while driving a screw, but that these fractures resulted in no procedural delays or impact to the patient.It was reported that the tips had become twisted and compromised over use and time.No additional instrumentation was required to complete the surgery and all pieces were removed prior to rod insertion.This is report one of three for this event.
 
Additional Manufacturer Narrative
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in g1 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip has twisted and fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference reports 3012447612-2024-00152 through 3012447612-2024-00154.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD CANNULATED DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key19716308
Report Number3012447612-2024-00152
Device Sequence Number8290163
Product Code HXX
UDI-Device Identifier00887868273681
UDI-Public(01)00887868273681(10)AD76
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number730M0060
Device Lot NumberAD76
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/30/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/10/2024
Supplement Report FDA Received Date12/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/09/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-