• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE - I10 SLIM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE - I10 SLIM Back to Search Results
Model Number EG27-I10
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the operation channel (primary) stuck accessory/object.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the operation channel (primary).In addition, our technician confirmed that the ccd drive pcb fluid damage, the electrical pin connector corroded, the lg connector corroded, and the biopsy inlet t-piece stuck object inside t-piece; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.This defect occurred not during procedure.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is not during procedure.There was no report of patient harm.Operation/suction channel clogged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
VIDEO GASTROSCOPE - I10 SLIM
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key19717375
Report Number9610877-2024-55279
Device Sequence Number12798763
Product Code FDS
UDI-Device Identifier04961333186846
UDI-Public04961333186846
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K131902
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG27-I10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2024
Initial Report FDA Received Date07/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-