Catalog Number 222002
Medical Device Problem Codes Break (1069); Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the device is shown in used condition.Biological matter can be observed over the device.The anchor was found broken in half.The overall complaint was confirmed as the observed condition of the 4.5hlx adv br anc-dyna 3-sut would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to procedural variables, such handling of the device or product interaction during procedure; as per ifu, inserting the awl or drill to less than the specified depth, axial misalignment, or levering with the anchor upon insertion may result in anchor fracture.Do not apply a bending force to the inserter.This can damage the anchor or inserter tip, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: concomitant med products and therapy dates: healix advance br 3 suture anchor w/dynacord device, (b)(6) 2024.E3: reporter is a j&j sales representative.The device manufacture date is unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
This is report 1 of 2 for (b)(6) it was reported by the sales rep via phone that during a rotator cuff repair procedure on (b)(6) 2024 two 4.5mm healix advance br anchor with dynacord suture (blue, white/blue/green), with needles was broken upon insertion and tips cracked.Another like device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5HLX ADV BR ANC-DYNA 3-SUT
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 
3035526892
MDR Report Key19718002
Report Number1221934-2024-02369
Device Sequence Number18175327
Product Code MAI
UDI-Device Identifier10886705029396
UDI-Public(01)10886705029396(10)UCXADX(17)270228
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K173859
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number222002
Device Lot NumberUCXADX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2024
Supplement Date Received by Manufacturer07/11/2024
Initial Report FDA Received Date07/11/2024
Supplement Report FDA Received Date07/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-