| Catalog Number |
153509 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the customer received defective drain bags where the tube went into the bag there was a bend in it which was preventing it from working properly.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿incorrect assembly operation / components placement / accessories (incorrect assembly)".However, there was insufficient information to confirm this potential root cause.The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer received defective drain bags where the tube went into the bag there was a bend in it which was preventing it from working properly.
|
| |
|