| Model Number |
A22005A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the telescope had a broken eyepiece.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.E1/establishment name: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: e1(unmark company representative).Additional information added to field h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's allegation was confirmed.Based on the results of the investigation, it is likely that the event was caused by improper handling and the application of excessive force.Olympus will continue to monitor field performance for this device.
|
| |
|