• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-450-14
Medical Device Problem Codes Stretched (1601); Material Deformation (2976); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that seven pipelines failed to open distally and became stretched.The patient was undergoing treatment for an unruptured, saccular aneurysm located in the right posterior communicating artery.The max diameter was 5mm, and the neck diameter was 3mm.The landing zone was 4mm distal and 5mm proximal.The patient's vessel tortuosity was severe. dual antiplatelet treatment was administered, and the pru level was said to be 80.It was reported that the physician tried with multiple pipelines in various sizes, but the severe tortuosity was causing the devices to stretch and be unable to deploy to the physician's standard.They were finally able to open a pipeline distally and drag it into place, allowing for the device to open and treat the patient successfully.The entirety of the pipelines had been placed in a bend when they failed to open.The pipelines had been resheathed more than 2 times.No additional steps were taken to open the devices.More than 50% of the device had been deployed when the last 4 pipelines failed to open. the patient did not experience any injury or complications.Angiographic results post procedure showed that the patient was treated successfully.The devices were prepared according to the instructions for use (ifu).Ancillary devices include a cook shuttle guide catheter and phenom 27 microcatheter.
 
Additional Manufacturer Narrative
Continuation of d10: product id ped2-500-30 (b369079); product id ped2-500-12 (b350774); product id ped2-500-14 (b408422); product type: ; implant date ; explant date product id ped2-500-16 (b324517); product id ped2-500-18 (b609558); product id ped2-500-20 (b266168); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key19721785
Report Number2029214-2024-01211
Device Sequence Number20884752
Product Code OUT
UDI-Device Identifier00763000284510
UDI-Public00763000284510
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2026
Device Model NumberPED2-450-14
Device Catalogue NumberPED2-450-14
Device Lot NumberB618823
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2024
Initial Report FDA Received Date07/11/2024
Date Device Manufactured09/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Patient SexMale
-
-