BAYER MEDICAL CARE INC STELLANT DUAL CT INJECTOR; CT INJECTION SYSTEM Back to Search Results
Model Number 85577026
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Air Embolism (1697)
Date of Event 06/10/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A system service check of the medrad® stellant flex injection system, serial number (b)(6), was performed on (b)(6) 2024, which confirmed that the injector was operating within bayer specifications.The disposables that were in use during the procedure were discarded by the site; therefore, they are not available for evaluation.The customer was unable to provide the lot number of the disposables used during the procedure; therefore, testing of retained samples was not possible.The offer of additional clinical applications training has been accepted by the customer and was provided by the clinical performance center.The medrad® stellant flex ct injection system operation manual cautions the user as follows: warning: air embolization can cause death or serious injury to the patient.Do not connect a patient to the injector until all trapped air has been cleared from the syringe and fluid path.This information does not constitute an admission that the device, the company, or its employees caused or contributed to a reportable event.
 
Event or Problem Description
Bayer medical care inc.Was informed that a 54-year-old female undergoing a ct scan with contrast suffered an alleged air injection while connected to a medrad® stellant flex injection system (sn (b)(6)).It was reported that an estimated 15-20 ml of air was noticed on the displayed images.The patient was subsequently admitted to the hospital from the emergency department, and after an undisclosed period of time, the patient was discharged to home with no further issues reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT DUAL CT INJECTOR
Common Device Name
CT INJECTION SYSTEM
Manufacturer (Section D)
BAYER MEDICAL CARE INC
1 bayer drive
indianola PA 15051
Manufacturer (Section G)
BAYER MEDICAL CARE INC
1 bayer drive
indianola PA 15051
Manufacturer Contact
jason johnson
1 bayer drive
indianola, PA 15051
4124063016
MDR Report Key19721943
Report Number2520313-2024-00015
Device Sequence Number18211537
Product Code DXT
UDI-Device Identifier00616258011219
UDI-Public(01)00616258011219
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K192370
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number85577026
Device Catalogue NumberSCT FLEX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/12/2024
Initial Report FDA Received Date07/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient SexFemale
-
-