ARTHREX, INC. SCORPION NEEDLE, KNEE; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number SCORPION NEEDLE, KNEE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a knee surgery it was found that the scorpion was blocked as prior to this event a needle broke off and remained stuck inside the device.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update 26-jun-2024 dw: further information were provided that the needle broke during the same surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
The complaint allegation was confirmed.One unpacked ar-12990n-1 serial/batch number (b)(6) was received for evaluation.The needled was received broken inside the ar-12990, batch 10474070.Visual evaluation of the ar-12990, batch 10474070 noted obstruction on the suture slot at the tip of the instrument, which is presumable to be the tip of the needle.No functional testing was performed because of a broken device inside the instrument.The most likely cause for the reported failure is a user error.Per dfu-03920-sub-eo, "to help avoid needle breakage and potential patient injury: do not resterilize or reuse the scorpion¿ suture passer needle.Avoid striking bone or using excessive force to pass the needle through fibrous/calcific tissue.If excessive force is encountered, replace the needle, reposition the device, and pass in a different area.Ensure a secure grasp of tissue to prevent the needle from buckling under the tissue.Avoid ¿dry,¿ repetitive actuations outside the surgical site.Abrasion of the needle surface can occur that may inhibit proper function.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCORPION NEEDLE, KNEE
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19722311
Report Number1220246-2024-06708
Device Sequence Number18194595
Product Code LXH
UDI-Device Identifier00888867266520
UDI-Public00888867266520
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCORPION NEEDLE, KNEE
Device Catalogue NumberAR-12990N-1
Device Lot Number15101671
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2024
Supplement Date Received by Manufacturer09/16/2024
Initial Report FDA Received Date07/11/2024
Supplement Report FDA Received Date09/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-