The complaint allegation was confirmed.One unpacked ar-12990n-1 serial/batch number (b)(6) was received for evaluation.The needled was received broken inside the ar-12990, batch 10474070.Visual evaluation of the ar-12990, batch 10474070 noted obstruction on the suture slot at the tip of the instrument, which is presumable to be the tip of the needle.No functional testing was performed because of a broken device inside the instrument.The most likely cause for the reported failure is a user error.Per dfu-03920-sub-eo, "to help avoid needle breakage and potential patient injury: do not resterilize or reuse the scorpion¿ suture passer needle.Avoid striking bone or using excessive force to pass the needle through fibrous/calcific tissue.If excessive force is encountered, replace the needle, reposition the device, and pass in a different area.Ensure a secure grasp of tissue to prevent the needle from buckling under the tissue.Avoid ¿dry,¿ repetitive actuations outside the surgical site.Abrasion of the needle surface can occur that may inhibit proper function.".
|