| Catalog Number |
TSVM6B8 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
|
| Date of Event |
05/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported by phone that he implant in site 31 was removed due to peri-implantitis.Patient stated she had been having pain around implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint (b)(4).A4: patient weight is not provided/unknown.E1: initial reporter¿s title is not provided/unknown.G4: additional 510(k) number is k101880.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b3.Date of event.B4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Dhr review was completed for the subject lot number 1260352.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 1260352 for similar events and one (1) other complaint was identified.Investigation has yet to be completed and results are unavailable at this time.Review completed utilizing keywords: peri-implantitis.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text: summary investigation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|