MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL AMT SZ8 125 STD COL; CORAIL IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL AMT SZ8 125 STD COL; CORAIL IMPLANTS : HIP FEMORAL STEM Back to Search Results
Catalog Number L971208
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
The stem inserter became stuck in the definitive stem.It wouldn't come out of the stem's tine when the surgeon tried to remove it and the entire definitive stem ended up being pulled out in the process.An osteotome was also used to try and remove the stem from the stem inserter.There was a burr which was identified on the end of the stem inserter after the case, it is unknown whether this was caused by the osteotome or if it was already there and was causing the stem inserter to get stuck.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: a3a, b7.Corrected (g3) alert date.The correct alert date for this complaint is 7/02/2024 based on aei note a-11416016 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "the stem inserter became stuck in the definitive stem.It wouldn't come out of the stem's tine when the surgeon tried to remove it and the entire definitive stem ended up being pulled out in the process.An osteotome was also used to try and remove the stem from the stem inserter.There was a burr which was identified on the end of the stem inserter after the case, it is unknown whether this was caused by the osteotome or if it was already there and was causing the stem inserter to get stuck." the product was not returned to depuy synthes; however, photos were provided for review.See attachment (b)(4).The investigation was not able to confirm the reported event as the complaint condition was represented clearly in the provided photo.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.
 
Event or Problem Description
Additional information received states that the procedure was completed successfully.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL AMT SZ8 125 STD COL
Common Device Name
CORAIL IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex IN 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR   69801
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19722332
Report Number1818910-2024-15051
Device Sequence Number8130504
Product Code MEH
UDI-Device Identifier10603295455387
UDI-Public(01)10603295455387(17)250430(10)9506813
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K192946
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue NumberL971208
Device Lot Number9506813
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2024
Supplement Date Received by Manufacturer07/20/2024
08/06/2024
10/22/2024
Initial Report FDA Received Date07/11/2024
Supplement Report FDA Received Date07/22/2024
08/07/2024
10/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 NEUT 32IDX48OD.; APEX HOLE ELIM POSITIVE STOP.; DELTA CER HEAD 12/14 32MM +5.; PINN 100 W/GRIPTION 48MM.
Patient SexFemale
-
-