SYNTHES GMBH 1.8MM DRILL BIT/QC/125MM; DRILL, BONE, POWERED Back to Search Results
Catalog Number 310.520
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2024 during a orthopedic surgery, the drill bit did not fit into the quick coupling adapter for the power equipment.This report is for one (1) 1.8mm drill bit/qc/125mm.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: photo investigation: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the device was inside primary packaging.However, manufacturer confirmed parts from this lot to be oversized due to a worn out insert causing a small burr, and buffing not performed properly, therefore, the reported event can be confirmed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the 1.8mm drill bit/qc/125mm would contribute to the complained device issue.Based on the information currently available, a product issue was identified during the investigation.This product issue will be addressed through depuy synthes quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Physical device investigation: the product was returned to depuy synthes for evaluation.A dimensional inspection was performed by the manufacturer and confirmed parts to be oversized due to a worn out insert causing a small burr, and buffing not performed properly.Therefore, an issue with correct assembling can be confirmed.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the 1.8mm drill bit/qc/125mm would contribute to the complained device issue.Based on the information currently available, a product issue was identified during the investigation.This product issue will be addressed through depuy synthes quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Relevant actions have been taken to address the complaint issue.Device history review (dhr): part # 310.520, synthes lot # u432724, supplier lot # u432724, release to warehouse date:16 jun 2023, supplier: (b)(4).No ncr's generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.8MM DRILL BIT/QC/125MM
Common Device Name
DRILL, BONE, POWERED
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key19722469
Report Number8030965-2024-08566
Device Sequence Number15522615
Product Code DZI
UDI-Device Identifier10887587035192
UDI-Public(01)10887587035192(11)230616(10)U432724
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K043310
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number310.520
Device Lot NumberU432724
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2024
Supplement Date Received by Manufacturer08/08/2024
Initial Report FDA Received Date07/11/2024
Supplement Report FDA Received Date08/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1.8MM DRILL BIT/QC/125MM
Patient SexUnknown
-
-