Model Number 5076-58
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Capturing Problem (2891)
Health Effect - Clinical Codes Non specific EKG/ECG Changes (1817); Syncope/Fainting (4411)
Date of Event 08/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: w1dr01 ipg implanted on (b)(6) 2018.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient experienced syncope and pause was noted while transporting to hospital.Right ventricular (rv) lead exhibited high and undefined lead impedance warning in unipolar and bipolar configuration and polarity switch.Recent increase in thresholds were noted on the rv lead.The right atrial (ra) lead exhibited out of range high unipolar impedance.Over-sensing was noted on the rv lead during non-sustained ventricular tachycardia (vt) episodes.The leads were capped and replaced.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS LEAD MRI SURESCAN
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112
7635269277
MDR Report Key19728185
Report Number3008973940-2024-05035
Device Sequence Number18215445
Product Code DTB
UDI-Device Identifier00763000014704
UDI-Public00763000014704
Combination Product (Y/N)Y
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/13/2019
Device Model Number5076-58
Device Catalogue Number5076-58
Was Device Available for Evaluation? No
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 06/17/2024
Initial Report FDA Received Date07/12/2024
Date Device Manufactured12/02/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age79 YR
Patient SexMale
-
-