| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report from a baxter technician stating the code calls were not alerting at the secondary location.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the rcbs needed to be reeboted.The navicare® nurse call¿ (nnc) system is a comprehensive communication and information management system that helps caregivers provide the best possible care to patients in the most efficient manner.The system facilitates communication within a nursing unit.Caregivers have the ability to quickly locate and communicate with a fellow staff member, and patients can easily communicate with caregivers, the nursing station, or directly to their primary caregiver.The hrc technician arranged to reboot of rcbs connected to code switch relays for the customer to resolve the issue.Although there was no injury associated with this event, if the reported issue were to recur it could lead to injury or death.Therefore hillrom is reporting this complaint.
|
| |
|