|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The microsensor (id 826633) was returned for evaluation.Device history record (dhr) - reported lot number 6250936, and no anomalies were identified during the manufacture or packaging of the device.However, the returned device for failure analysis doesn¿t match the reported lot number.Therefore, we have reviewed the returned device¿s lot number 6345637 and no anomalies were identified during the manufacture or packaging of the device.Failure analysis - the catheter and sensor were inspected.No defects were observed.The fittings were inspected.The blue cap was fully tightened.A syringe with water was used to introduce fluid to the catheter.Water was observed leaking from the blue cap.Based on the evaluation, the complaint is confirmed.Root cause analysis - per the complaint background, csf leak.The complaint was confirmed during investigation and failure analysis evaluation.A potential root cause is improper cap tolerance or damage to the luer-lock connector.
|