• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number CVPLP2000
Medical Device Problem Codes Peeled/Delaminated (1454); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 06/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: e1455b, e1455b edge ins shrouded blade elecx25 (lot#:240500132r), vlft10gen, vlft10gen ft series energy platformx1 (serial#:(b)(6)), se3690, se3690 rapidvac smoke evacuator 110vx1 (serial#:(b)(6)), e7507, e7507 polyhesive ii rem ret el w/cord (lot#:unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during left axillary lymph node dissection, during dissection of tissue within surgical site, the electrode with the safety sleeve was pulled for a case.The scrub tech when realizing it was not compatible with the cvplp2000, attempted to cut the safety sleeve off, and then forced the electrode onto the bovie pencil with a pair of graspers or similar device and that tore the insulated lining on the bovie pencil, causing an alternate pathway for the electricity which caused the patient small 1cm burn inside the surgical dissection site up to the edge of the incision.Items listed were replaced and items involved were sequestered in order to complete the case.The burn has been bandaged, and the patient is doing well with no concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19735793
Report Number2182208-2024-02805
Device Sequence Number8375804
Product Code GEI
UDI-Device Identifier10884521834613
UDI-Public10884521834613
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K103375
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2026
Device Model NumberCVPLP2000
Device Catalogue NumberCVPLP2000
Device Lot Number230806164X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2024
Initial Report FDA Received Date07/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-