H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation; the stent graft was found partially deployed and the outer sheath was fractured which leads to confirmed results.A provided photo shows the distal part of a bloody stent graft delivery system with the stent graft partially deployed.It was reported that a 0.035", the tracking vessel was neither tortuous nor calcified, the lesion was pre-dilated and the device was flushed prior to use.The two-way stopcock was detached and missing from the sample and this detachment is considered not to have a bearing on stent graft deployment.Based on provided photos and the evaluation of the returned sample, the investigation is closed with confirmed results for sheath fracture and partial deployment.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.The instructions for use state: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device'.Regarding preparation of the device the instructions for use state that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment".Regarding accessories the instructions for use states: "prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location.The use of an appropriately sized introducer sheath is recommended"; the labeling pictograms indicate an introducer size of 9f.H10: d4 (expiration date: 06/2025) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|