• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC I; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC I; INTRAOCULAR LENS Back to Search Results
Lot Number SB19733
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Inflammation (1932); Visual Impairment (2138); Eye Pain (4467)
Date of Event 06/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There is no evidence to suggest that any aspect of the device was responsible for the reported event.Additional information is being sought and assessment by the medical monitor is ongoing, once completed a supplemental report will be submitted.
 
Event or Problem Description
Lenstec received an email staing "visual acuity decline, pupillary exudation".
 
Additional Manufacturer Narrative
Based on the review of the batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the devices were conducted correctly.The endotoxin results were within acceptable limits, and we have not received any other previous complaints from these batches.The lenses that were returned for testing were intact and no defects were noted.The saline solution was free of particles.Additionally, lenstec is unable to comment on the compatibility of the injector reported as used for the surgery amongst other surgical instruments or the sterility processes used by the facility.Furthermore, lenstec's medical monitor stated that there was no indication that anything was abnormal with any of these implants.There was no evidence to suggest that any aspect of these devices was responsible for the reported incident.Furthermore, lenstec confirms that the lenses, their design and the manufacturing process are not at fault due to the extensive testing that we have performed on this model.
 
Event or Problem Description
Lenstec received an email stating "visual acuity decline, pupillary exudation".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC I
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
Manufacturer (Section G)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB  
Manufacturer Contact
jimmy chacko
1765 commerce ave. n.
st. petersburg, FL 33716
7275712272
MDR Report Key19737093
Report Number9613160-2024-00051
Device Sequence Number12539617
Product Code HQL
UDI-Device Identifier00844369029506
UDI-Public00844369029506
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Lot NumberSB19733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2024
Supplement Date Received by Manufacturer06/13/2024
Initial Report FDA Received Date07/13/2024
Supplement Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-