Unknown lot number: d.4.Medical device expiration date: unknown h.4.Device manufacture date: unknown the information for the additional 510k is as follows: g5.Pma / 510(k)#: k020322, k021954, k022172, k023273, k023301, k024152, k030677, k031306, k031679, k031679, k032131, k033784, k033907, k040006, k040106, k040716, k050555, k051689, k053241, k060214, k060217, k060218, k060493, k082538, k082852, k082913, and k131331.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Investigation summary this complaint is for no mic results for trimethoprim / sulfamethoxazole (sxt) when using phoenix panel pmic-106 (catalog number 448416) batch number unknown.The customer did not provide isolates, or panel returns but provided phoenix generated lab reports for the investigation.The batch number was not provided; therefore, this complaint is unconfirmed.A review of quality notifications could not be performed since a batch number was not provided.A review of complaints could not be performed since a batch number was not provided.Complaint trending was performed, and no trends were identified associated with this defect.Bd will continue to monitor for trends and take action as necessary.Please continue to communicate any additional concerns.
|