|
This supplemental report is being submitted to provide the results of the final investigation.Updates added to the following fields: d9, h3, h6 the device was returned to olympus for inspection, and the reported failure was confirmed.Based on the damage described, the damage was likely caused by thermal and mechanical factors.It cannot be determined whether the insulating insert was damaged or worn prior to the incident, or whether the damage occurred during the last preparation (cleaning, disinfecting, sterilization) of the instrument or during the last procedure.Based on the results of the investigation, it is likely the following led to the malfunction: wear and tear and wrong handling, such as a fall, overstress, or blow.There are also cracks on the insulation material that are hard to spot during a visual inspection.Olympus will continue to monitor field performance for this device.
|