| Model Number |
M00545150 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a autotome rx 49 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with choledocholithotomy procedure performed on (b)(6) 2024.During the procedure, while cutting the papilla, the cutting wire suddenly got fractured.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: a photo of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
Block a2: patient's exact age is unknown; however, it was reported that the patient was over the age of 18.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a autotome rx 49 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with choledocholithotomy procedure performed on (b)(6) 2024.During the procedure, while cutting the papilla, the cutting wire suddenly got fractured.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: a photo of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
Block a2: patient's exact age is unknown; however, it was reported that the patient was over the age of 18.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: investigation results the returned autotome rx 49 was analyzed, and a visual and microscopic inspection found that the cutting wire was blackened, kinked and broken from the proximal pierce hole.The customer provided a picture where was possible to observe that the cutting wire was broken and kinked.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of cutting wire break was confirmed.The results of the analysis performed on the returned device found that the cutting wire was kinked and broken from the proximal pierce hole.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.The break in the cutting wire could have been generated if there was contact between the device and the scope during energization or if the device exceeded the maximum of voltage during procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wires integrity, causing premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can bend the cutting wire as observed in the product analysis.The investigation findings and all information available concludes the most probable root cause is an adverse event related to procedure.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
|
| |
|