Corrected information is provided in section d.4.Additional information is provided in sections e.1, d.9, h.3, h.6 and h.11.Three opened trocar assemblies, in a surgical micro-puncture trocar (smpt), in a bag were received for the report of fluid leakage was observed from the valve during the procedure.Samples were visually inspected and found to be conforming.Functional leak testing was performed and all three samples were found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned samples were found to be visually and functionally conforming, therefore fluid leakage was observed from the valve during the procedure as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the trocars were manufactured to specifications.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|