• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 3MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 3MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number ILPC343U
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).
 
Event or Problem Description
Doctor reported that the screw abutment broke inside the implant.Doctor confirmed that the implant is fine.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report, b5: describe event or problem, g3: date received by manufacturer, g6: type of report, h1: type of reportable event, h2: follow up type, h3: device evaluated by manufacturer, h6: adverse event problem and h10: additional narrative.Zimvie received one (1) ilpc343u, (certain low profile one-piece abutment 3d.4mm(d) x 3mm(h)) for evaluation.Visual evaluation was performed, a fracture has been identified at the threads.Device history record (dhr) review was completed for the subject lot number 2017080833.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2017080833 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : screw based on the investigation and risk management file review, the most likely root cause determined from the investigation was clinician incorrectly engages abutment screw into implant.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the threads.The reported event is confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 3MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key19739070
Report Number0001038806-2024-01479
Device Sequence Number18072581
Product Code NHA
UDI-Device Identifier00844868021094
UDI-Public(01)00844868021094(17)220912(10)2017080833
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K092341
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/11/2022
Device Catalogue NumberILPC343U
Device Lot Number2017080833
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/15/2024
Supplement Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-