ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 25CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
NSLX125C |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 7/15/2024.D4 batch #: unknown.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.D4: udi: the manufacturing date is currently not available.Therefore, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that, during open gastrectomy, in the middle of the surgery, it was noticed that the tip of the shaft was broken.The doctor said that he did not put effort forcibly.There were no pieces left in the patient.There is a possibility of the reason of this case that a nurse did not check precautions properly.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 8/5/2024; d4 batch #: a9ea63.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was returned with the distal guide weld broken.No top jaw missing was noted.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.The damage to the distal guide was not enough to prevent the jaws to open and close as expected.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached on the cause of the reported event.Please refer to the ifu for proper handling of the device, note that ¿caution: do not grasp tissue beyond the electrode surface, in the hinge of the jaws.Do not overfill the jaws of the instrument with tissue.This could result in difficulty opening the jaws, partially cutting tissue, and unintended injury.
|
| |
|