MAUDE Adverse Event Report: ARTHREX, INC. BURR, ROUND, 8 FLUTE 5.5MM X 13CM; BURR,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BURR, ROUND, 8 FLUTE 5.5MM X 13CM; BURR,SURGICAL Back to Search Results
Model Number BURR, ROUND, 8 FLUTE 5.5MM X 13CM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 06/26/2024, it was reported by a facility representative via sems (b)(4) that two (qty.2) ar-8550rbe round burr attachments were broken in the hub at the attachment connection.This occurred during use in a case with no patient impact.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Complaint allegation is confirmed.One unpackaged ar-8550rbe serial/batch number (b)(6) was received for investigation.Functional testing could not be conducted because the connector hub was damaged internally.A visual inspection revealed that the connector hub is broken internally.The most likely cause of the reported failure is the application of excessive force during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, ROUND, 8 FLUTE 5.5MM X 13CM
Common Device Name
BURR,SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19740051
Report Number1220246-2024-06736
Device Sequence Number7785636
Product Code GFF
UDI-Device Identifier00888867044074
UDI-Public00888867044074
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBURR, ROUND, 8 FLUTE 5.5MM X 13CM
Device Catalogue NumberAR-8550RBE
Device Lot Number15213867
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2024
Supplement Date Received by Manufacturer09/18/2024
Initial Report FDA Received Date07/15/2024
Supplement Report FDA Received Date09/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-