On 04/24/2024: it was reported, by a patient via phone.That after, undergoing a distal femur osteotomy on (b)(6) 2023, one of the six implanted screws broke.This was discovered 3 1/2 months post-op in an x-ray.On (b)(6) 2023, when this procedure took place, an ar-13280-60, cancelleous screw; an ar-13380-54, cortical screw; an ar-13280-60, cancelleous screw; an ar-13380-50, cortical screw; an ar-13280-65, cancelleous screw and an ar-13380-46, cancelleous screw were implanted.It has not been confirmed, which screw broke.A post-op x-ray was taken 2 months, post-op where everything looked good.However, the x-ray taken 3 1/2 months post, showed a broken screw that had partially made its way.The integrity of the healing bone will determine, when a revision will be taking place.Additional information, was provided by the patient.Where he stated, that a test had been run to confirm, there were no deficiencies with his health, that would cause this failure.The patient will be undergoing a revision surgery with a trauma surgeon.There patient stated, his femur is now short on both sides, due to the breakage.Additional information, was provided on 06/27/2024.Where the patient reported, that a revision procedure, had taken place on (b)(6) 2024.They removed all of the screws and bracket.Then, it was replaced with other hardware.As well as, a femur rod added.Per the surgeon who removed the devices, it was the 6.5mm x 60mm, that was broken.
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to over-torquing the screw during insertion and/or misaligned insertion of the screws; causing pressure to be applied to the shaft of the screw while implanted.
|