| Model Number |
SCREW, CANCELLOUS 6.5 X 60MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 04/24/2024, it was reported by a patient via phone that after undergoing a distal femur osteotomy on (b)(6) 2023 one of the six implanted screws broke.This was discovered 3 1/2 months post-op in an x-ray.On (b)(6) 2023, when this procedure took place, an ar-13280-60 cancellous screw, an ar-13380-54 cortical screw, an ar-13280-60 cancelleous screw, an ar-13380-50 cortical screw, an ar-13280-65 cancelleous screw, and an ar-13380-46 cancelleous screw were implanted.It has not been confirmed which screw broke.A post-op x-ray was taken 2 months post-op where everything looked good; however, the x-ray taken 3 1/2 months post showed a broken screw that had partially made its way.The integrity of the healing bone will determine when a revision will be taking place.Additional information was provided by the patient where he stated that a test had been run to confirm there were no deficiencies with his health that would cause this failure.The patient will be undergoing a revision surgery with a trauma surgeon.There patient stated his femur is now short on both sides due to the breakage.Additional information was provided on 06/27/2024, where the patient reported that a revision procedure had taken place on (b)(6) 2024 they removed all of the screws and bracket.Then, it was replaced with other hardware.As well as a femur rod added.Per the surgeon who removed the devices, it was the 6.5mm x 60mm that was broken.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to over-torquing the screw during insertion and/or misaligned insertion of the screws; causing pressure to be applied to the shaft of the screw while implanted.Complaint allegation is confirmed.See photos attached.
|
| |
|
Search Alerts/Recalls
|
|
|