Catalog Number 121936058
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
This is the second revision of this acetabulum, the first procedure was reported with (b)(4).The polyethylene liner became dissociated from the pinnacle sector cup in that instance, and the same has happened here.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).D4: udi: as the catalog/model number was not provided, the (01)gtin is not available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: this is the second revision of this acetabulum; the first procedure was reported with (b)(4).The polyethylene liner became dissociated from the pinnacle sector cup in that instance, and the same has happened here.There is thought that there may be something wrong with the interface between the liner and the cup.I have requested that the liner and cup be decontaminated and available to be sent for inspection.The sector cup was explanted and revised with a pinnacle multihole cup and 40mm liner.The product was returned to depuy synthes for evaluation.Visual analysis of the returned device revealed that the rim of the pinn mar neut 36idx58od has (b)(4) anti-rotation device (ard) tabs sheared off.The broken fragments were not returned for evaluation.Additionally, the liner appeared to have been edge loaded causing eccentric deformation in the rim of the device and ultimately contributing to the failure of the anti-rotation devices (ards).Based on the observations of the pinn mar neut 36idx58od and the returned femoral head, it is reasonable to conclude that a disassociation event occurred between the liner and the cup.There is no indication that a design or manufacturing issue has caused or contributed to the reported event.Although the liner failure cannot be traced to design or manufacturing, a definite root cause cannot be established due to there being multiple factors that may influence eccentric loading.Consideration must be given to all potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.It is highly recommended to the patient to consult with their healthcare professional for further assessment.A manufacturing record evaluation was performed for the finished device [121936058/4182463] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.A dimensional inspection for the liner was unable to be performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the pinn mar neut 36idx58od would have contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to component failure, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device [121936058/4182463] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.Device history review: a manufacturing record evaluation was performed for the finished device [121936058/4182463] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 36IDX58OD
Common Device Name
HIP ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19742684
Report Number1818910-2024-15178
Device Sequence Number7719775
Product Code LPH
UDI-Device Identifier10603295014478
UDI-Public(01)10603295014478(17)280531(10)4182463
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K010171
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121936058
Device Lot Number4182463
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2024
Supplement Date Received by Manufacturer08/07/2024
09/24/2024
Initial Report FDA Received Date07/15/2024
Supplement Report FDA Received Date08/12/2024
09/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DLT TS CER HD 12/14 36MM +8.5.; PINN SECTOR W/GRIPTION 58MM.; UNK HIP FEMORAL HEAD CERAMIC.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-