| Catalog Number |
CDH23P |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic esophagectomy, used during anastomosis of esophagus and stomach.The device could not be fired with pulling the firing knob.No electrocautery was used immediately prior to the procedure.There was nothing out of the ordinary including the setting.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 7/15/2024.D4 batch # unk.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 8/15/2024.D4: batch #787a33.Investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the cdh23p device arrived with no apparent damage.The breakaway washer was present and uncut, the device was fully loaded with staples, indicating that the device had not been fired.However, when the test battery was installed the green checkmark illuminated indicating that the device was not reset.No functional test was performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the cam was rotated so that it was touching the stop switch actuator.In addition, a tyvek was received along with the device.Additionally a photo was provided at product complaint level and it shows the device with the battery inserted, also the anvil was not placed into the trocar, the safety disengaged and the firing trigger can be seen disengaged.This defect has been correlated to a manufacturing process.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.A manufacturing record evaluation was performed for the finished device batch number 787a33, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|