MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ACTIVE ARTICULATION G7 DUAL MOBILITY E1 HIP BRG 22X36MM; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ACTIVE ARTICULATION G7 DUAL MOBILITY E1 HIP BRG 22X36MM; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Subluxation (4525)
Date of Event 05/14/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision procedure 1 month and 20 days post implantation due to intra-prosthetic luxation.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).D10: 00801802220 item name femoral head sterile product do not resterilize 12/14 taper, lot # 66225522.G2: foreign: belgium.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4 g3 g6 h2 h6 h11 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no related deviations or anomalies during manufacturing.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues: left total hip arthroplasty with superior and likely posterior dislocation as well as status post reduction on the fourth image.A definitive root cause cannot be determined.Event is confirmed via medical records if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVE ARTICULATION G7 DUAL MOBILITY E1 HIP BRG 22X36MM
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19744018
Report Number0001825034-2024-01835
Device Sequence Number20779731
Product Code LPH
UDI-Device Identifier00880304527393
UDI-Public(01)00880304527393(17)270716(10)685040
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K161190
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number010000991
Device Lot Number685040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2024
Supplement Date Received by Manufacturer12/04/2024
Initial Report FDA Received Date07/15/2024
Supplement Report FDA Received Date12/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-