• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER ADVANCED DSP TIP FORCEPS; FORCEPS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER ADVANCED DSP TIP FORCEPS; FORCEPS, OPHTHALMIC Back to Search Results
Catalog Number 725.44P
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A nurse reported that two ophthalmic surgical forceps could not be opened after being closed during vitrectomy surgery.The surgery was completed on the same day.No patient impact was reported.
 
Additional Manufacturer Narrative
A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A review of the database for the non-conformance events and the complaints showed that the given lot number was not concerned by any deviation and that there are no additional complaint associated with it.The investigation is completed based on the sample evaluation documented below.The sample was received in its outer blister, covered with the tyvek lid foil showing the lot information.The inner blister, which offers protection to the instrument during shipment, was not used.The sample shows no macroscopic signs of damage but it is evident that the forceps is in a closed state when the handle is not actuated, with the grasping arms being pulled far into the metal shaft.The sample exhibits some signs of surgery residues.The sample was visually inspected with the aid of a photomicroscope using various magnifications and was functionally tested.During the inspection, it was confirmed that the forceps gets stuck in a closed state and can only get opened again by manually pushing down the metal shaft, indicating the bonding connection between the shaft and the sleeve got broken.This confirms the customer¿s complaint.As the blister was opened by the customer, the product was handled.However, it is not possible to determine what situation could have caused the malfunction and no root cause can be identified conclusively.No root cause for the customer's reported event could be determined, since the situation that lead to the instrument failure cannot be reconstructed.The exact root cause for the customer's reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER ADVANCED DSP TIP FORCEPS
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19744299
Report Number3003398873-2024-00077
Device Sequence Number14344791
Product Code HNR
UDI-Device Identifier07612717071018
UDI-Public07612717071018
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue Number725.44P
Device Lot Number164JAE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2024
Supplement Date Received by Manufacturer08/30/2024
Initial Report FDA Received Date07/15/2024
Supplement Report FDA Received Date09/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-