| Catalog Number |
06407358190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The c702 module serial number was (b)(6).Two calibrations with 2 different reagent packs were performed on (b)(6) 2024; the signals were much higher than the previous calibration on (b)(6) 2024.Calibrations were performed again on (b)(6) 2024 with one of the same reagent packs and a new reagent pack.Qc was acceptable on (b)(6) 2024.Failing qc was also provided from (b)(6) 2024.There was no qc recovery data for (b)(6) 2024.No issues were identified during a review of the alarm trace data.The customer declined a service visit as the issue was resolved by replacing the reagent and recalibrating.The customer¿s troubleshooting actions resolved the issue.
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of qc issues and questioned high results for 35 patient samples tested for magnesium (mg2) on a cobas 8000 c 702 module.The customer removed the onboard reagent and recalibrated using the standby reagent.The following are examples of discrepant results for 2 patient samples.Patient 1 initial result was 3.6 mg/dl.The repeat result was 2.1 mg/dl.Patient 2 initial result was 3.4 mg/dl.The repeat result was 2.0 mg/dl.
|
| |
|
Search Alerts/Recalls
|